Safety ILS500 F
INFICON can not take any responsibility for the consequences arising from
inappropriate use of certain test pressures.
Warning
The ILS500 F must never be introduced to pressures higher than that approved for
the object to be tested and never beyond the ILS500 F specification.
Warning
Be sure to have a pressure relief valve in case of accidental tracer gas pressure
increase.
Warning
When dealing with high pressures, a blast protection is needed between the Test
Ports and the Test Object.
Warning
When dealing with test objects that cannot stand high pressure increase, make sure
to mount a flow control valve on the Test Ports.
Warning
Make sure not to confound Compressed Air and Tracer Gas.
Notice
The ILS500 F has no internal emergency stop circuit. ILS500 F is
prepared for integration into an external emergency stop circuit.
Check that all relevant legislation and safety standards are complied with
before putting the ILS500 F into service. See further information under
Installation.