4 status output, 5 tooling interface – INFICON Sensistor ILS500F Leak Detection System User Manual
Page 78

78
Technical Data
ninp69e1-a (1410)
11.4.4
Status Output
11.4.5
Tooling Interface
Connector:
6 pin male Weidmüller, Omnimate BL3.5. Mating screw terminal
included.
Purpose:
Test Status Outputs. Sourcing 24 VDC transistor outputs.
Pin
Signal
Type
Load
Comment
1
RUNNING
OUT
0.5 A
Cycle running.
2
ACCEPT
OUT
0.5 A
Tested part accepted.
3
REJECT
OUT
0.5 A
Tested part rejected.
4
ERROR
OUT
0.5 A
Summing error.
5
EOT/FILLED
OUT
0.5 A
End of test or gas filled indicator
(selectable).
6
COM
GND
-2.0 A
Common GND.
Notice
Gas filling status is available on the STATUS connector (pin 5) on the back
of the unit. Connect to a lamp for easy notification of “End of test” status.
Connector:
8 pin male Weidmüller, Omnimate BL3.5. Mating screw terminal
included.
Purpose:
Electrical tooling interface.