Tracer gas fill valve, External gas leaks – INFICON Sensistor ILS500F Leak Detection System User Manual

Page 55

background image

55

Troubleshooting

ninp69e1-a (1410)

Tracer Gas Fill Valve

This step tests that tracer gas fill valve opens to fill gas. Test will fail if tracer gas feed
pressure is too low. If this is the case, adjust pressure and restart hardware test from
beginning.

Possible results:

Valve works

Faulty!

Malfunction will result in:

Failed gas filling

External Gas Leaks

The ILS500 F is now prepared for a manual test for external leakage. Use a Leak
detector with hand probe to check for leakage.

1

Start by checking all connections between the ILS500 F and your test object.
Follow each test line carefully and check every joint.

2

Proceed to check around the gas valves and manifold inside the ILS500 F.

Hardware Test

CONTINUE

STOP

Vacuum:

Pressure:

0.00 bar

bar

Tracer Gas Fill Valve

Valve works!

0.50

Hardware Test

CONTINUE

STOP

Vacuum:

Pressure:

0.00 bar

bar

Check for Leaks with Hand Probe

0.50