Troubleshooting

52

ninp69e1-a (1410)

Evacuation Valve

Evacuation valve is checked for internal leakage.

Possible results:

No Internal Leakage

Internal Leakage

Internal leakage can result in:

False vacuum decay rejects

Increased tracer gas consumption

Venturi Pump

Checking max vacuum of Venturi pump.

Possible results:

Max Vacuum OK

Poor Max Vacuum

Poor max vacuum can result in:

Failed pre-evacuation

Slower evacuation

Hardware Test

CONTINUE

STOP

Vacuum:

Pressure:

-0.03

0.00

bar

bar

Evacuation Valve

No Internal Leakage

Hardware Test

CONTINUE

STOP

Vacuum:

Pressure:

-0.88

0.00

bar

bar

Venturi Pump

Max Vacuum OK