Bail Handle Compressed Gas Connectors

FasTest, Inc. 1646 Terrace Drive Roseville, MN 55113 Support: (800) 444-2373 Fax: (651) 645-7390 www.fastestinc.com

11

problem

recognized By

probable cause

recommended action

Connector’s thread
collets not expanding
properly during initial
hook-up to cylinder valve

Visual inspection of
connection joint

Short connection of connector
to valve

Visual inspection of valve.
Replace if damaged or worn.
Disconnect and reconnect
connector to valve. Be sure
actuator handle sleeve is fully
engaged. If problem is unresolved,
contact

FasTest

Loose connection

Connector is loose
despite proper
connection

Worn or damaged threads of
cylinder valve

Replace cylinder valve.

Damage, deformation
or distortion to connector
body, sleeve, and collet
threads. Possible internal
leakage

Visual inspection of
connector

Difficult operation of
connector

Improper operation

Remove connector from filling
operation immediately!
Return to

FasTest to determine

probable cause

Inability to connect to,
or a leakage with 540
and 580 RPV style
connectors and Residual
Pressure Valves

Inability to fully
actuate connector
actuator handle and/
or outer sleeve

(a) Bent actuator pin

(b) Damaged actuator piston

(a) Replace actuator pin

(b) Return to

FasTest for repair

Inability to connect or
leakage of RPV version
connector to non RPV
cylinder valves

Inability to fully
actuate and/or gas
leakage at initial filling

(c) Actuator pin not retracted

(c) Retract/remove actuator pin

according to specific connector
operation instructions