Det-Tronics R8471NH34001,2,3 Single Channel Gas Controller, NH3 User Manual
Page 23

21
95-8529
1.1
Section IV
System Maintenance
ROUTINE MAINTENANCE
OUTPUT DEVICES
All external equipment that is connected to the 
controller should be maintained according to the 
manufacturer’s recommendations.
CHECKOUT IN NORMAL MODE
The system should be checked periodically in the 
Normal mode to ensure that those items not checked 
by the controller diagnostic circuitry (such as output 
relays) are functioning properly.
CAUTION
Be sure to secure all output devices that are 
actuated by the system to prevent unwanted 
activation of this equipment, and remember 
to place these same output devices back into 
service when the checkout is complete.
SENSOR CELL REPLACEMENT
NOTE
Verify that no hazardous conditions exist or 
remove power from the sensor prior to replacing 
the sensor cell (unless an I.S. barrier is used with 
an intrinsically safe sensor).
To replace the sensor cell:
1. If the controller will remain powered during sensor
cell replacement, enter the sensor replacement 
mode before replacing the cell. 
NOTE
The sensor replacement mode inhibits all 
controller outputs to allow replacement of the 
sensor cell without removing power from the 
controller. In addition, this mode automatically 
sets the factory default values for sensor 
calibration. Alarm setpoints and calibration time 
are not affected.
CAUTION
Upon entering the Sensor Replacement mode, all 
previously entered sensor calibration information 
is lost. Sensor calibration must be performed, 
even if the sensor cell was not replaced.
To enter the Sensor Replacement mode, press and 
hold the Reset button for approximately 9 seconds 
until the digital display begins flashing and the CAL 
LED is illuminated. Release the Reset button.
2. Press the Set button. The FAULT LED also comes
on. The controller is now in the Sensor Replacement 
mode.
3. Remove the cap from the sensor base. See Figure
5. (There is no need to remove the sensor housing 
from the junction box.)
4. Remove and discard the old sensor cell. Check for
corrosion or contamination on the terminals on the 
sensor base, and clean if necessary.
5. Determine proper orientation for the new cell, then
carefully
plug it in.
NOTE
Handle the sensor cell carefully. To avoid 
possible damage, observe the normally accepted 
procedures for handling electrostatic sensitive 
devices. See form 75-1005 for additional 
information.
6. Be sure that the O-ring on the sensor housing is
in good condition, then place the cap back on the 
sensor base. Tighten only until snug. 
Do not over
tighten.
7. If power was removed prior to replacing the sensor
cell, re-apply power. Allow time for the unit to warm 
up and stabilize, then calibrate. 
NOTE
If power was removed from the controller during 
the sensor cell replacement procedure, the 
controller will automatically return to the Normal 
mode when power is restored (after a time 
delay). To prevent the possibility of alarms, allow 
the system to warm up in the sensor replacement 
mode. The sensor replacement mode can 
be entered during the power-up time delay by 
holding the Reset button for 7 seconds.
Press Reset to exit the Sensor Replacement mode. 
The controller will automatically enter the Normal 
mode. 
A Recommended Test Form is supplied at the back of 
this manual for recording maintenance performed on 
the system.
