INTRODUCTION

CARON’s 6540 series photostability chambers are designed specifically to meet ICH, FDA EMEA, and

Health Canada requirements for photostability testing according to ICH Q1B option II. UVA (black) and

VIS (cool white) lamps provide sequential light exposure.

The chamber conditions are conveniently displayed at eye-level and controlled through the touch-screen

display. Calibrated photon detectors measure light intensity. Lights can be manually turned on/off or

automatically based on exposure level (W-hr/m

2

or Lux-hr) or time. To enhance reliability, temperature and

humidity (optional) conditions are monitored and controlled throughout the test.

Specifications

Lamp Type, UVA Fluorescent, near UV (black lamp)

Lamp Type, VIS Fluorescent, cool white

Temperature Range 10 to 35

C

Temperature Control

0.2

C

Humidity Range (model 6545)

40 to 70% RH

Humidity Control (model 6545)

3%

Interior Dimensions 683 W x 610 D x 711 H mm (296 L; includes lamp space)

Features

Controls

● Icon based, easy to use, eye-level, touch-screen interface

● Automatically turns off lights at end of programmed exposure

● Alerts operator when lamps need changed

● Temperature and humidity deviation alarms

● Security lockout to prevent accidental tampering

Light Detectors

VIS and UVA, calibrated

Shelves

Two, sliding, 22 lbs (10kg) maximum load each

Construction, Interior

Stainless steel

Reflective Interior Surfaces

Specular aluminum

Chamber Footprint

955 W x 767 D mm

Chamber Height

1168 mm with leveling feet

Power

115V 60Hz, 208/230V 60Hz, or 230V 50Hz

Access Port

Two (44 W x 51 H mm each)

Door

Light tight, magnetic gasket seal

Room ambient conditions must be 20ºC to 25ºC, non-condensing. Exceeding 25ºC can result in operational

failure. The temperature and humidity ranges can be extended (optional). Humidity limited by 6ºC dew

point.

Models 6540-1 and 6545-1 were formerly models 6540A and 6545A respectively.

Models 6540-3 and 6545-3 were formerly models 6540E and 6545E respectively.