Caron 6545 User Manual
Page 44
6540 Series Operations Manual Rev_I.docx
4/7/2014
44
Calibration
The Calibration feature is used to adjust the light intensity, temperature and humidity (optional) against an
independent calibrated measuring device. A linear scale factor is set for both UVA and VIS light intensity.
An offset is set for temperature and humidity.
1. From the Setup Menu, press the
(CALIBRATION) button.
2. Press the
(UVA CALIBRATION) button.
3. Use the
(UP) and
(DOWN) arrow buttons to adjust the calibration scale factor so the light
intensity corresponds with an independent calibrated light sensor. The default scale factor is zero.
To increase the intensity, increase the scale factor. Scale factors can be negative.
4. Press
(EXIT) button to return to the calibration screen.
UVA light
calibration button
Humidity
calibration button
(optional)
Temperature
calibration button
VIS light
calibration button
UVA scale
factor
Light intensity
with calibration
scale factor
applied