Calibration

The Calibration feature is used to adjust the light intensity, temperature and humidity (optional) against an

independent calibrated measuring device. A linear scale factor is set for both UVA and VIS light intensity.

An offset is set for temperature and humidity.

1. From the Setup Menu, press the

(CALIBRATION) button.

2. Press the

(UVA CALIBRATION) button.

3. Use the

(UP) and

(DOWN) arrow buttons to adjust the calibration scale factor so the light

intensity corresponds with an independent calibrated light sensor. The default scale factor is zero.

To increase the intensity, increase the scale factor. Scale factors can be negative.

4. Press

(EXIT) button to return to the calibration screen.

UVA light

calibration button

Humidity

calibration button

(optional)

Temperature

calibration button

VIS light

calibration button

UVA scale

factor

Light intensity

with calibration

scale factor

applied