Alarms
There are high and low process alarms that can be set for UVA light, VIS light, temperature and humidity
(optional). Alarms automatically reset when the alarm condition is satisfied. Alarms may be silenced
(buzzer turns off) while the alarm condition continues. Process alarms are absolute, independent of the
corresponding set point. See maintenance section for recommended lamp hours.
The user set alarms are:
Maximum UVA lamp hours
UVA lights dim (minimum UVA light intensity)
Maximum VIS lamp hours
VIS lights dim (minimum VIS light intensity)
High temperature
Low temperature
High humidity
Low humidity
If the unit detects an error, it may alert the operator to additional alarms such as:
Lamps too hot on top bank*
Lamps too hot on bottom bank*
Top light bank not installed properly
Bottom light bank not installed properly
Humidity water not draining properly (humidity only)
Humidity water not filling properly (humidity only)
* This alarm requires manual resetting (acknowledge alarm).
To access the alarm screen, press the
(ALARM) button from the main.
Alarm message area
Set UVA lights, VIS lights, temperature
and humidity alarm buttons