beautypg.com

Caron 6545 User Manual

Page 53

background image

6540 Series Operations Manual Rev_I.docx

4/7/2014

53

Is there a temperature calibration factor applied?

Is there free air flow (no obstructions) next to all of the air vents on the front, right and back sides?

Unit reaches temperature set point but won’t stabilize

Are both access ports closed?

Has the door been open within 30 minutes?

Have the lights cycled on or off within 30 minutes?

If the chamber has humidity control, is it stable?

Unit won’t reach humidity set point

Are both access ports closed?

Is the humidity set point within the unit specification range?

Has the unit stabilized for 2 hours?

Is the ambient temperature between 20ºC-25ºC?

Is the drain draining? No kinks or up-hill traps in the line? Cycling power resets any silenced alarms.

After turning the unit on, wait 1 minute to see if ‘Water not draining properly’ alarm activates.

Is there a humidity calibration factor applied?

Is there free air flow (no obstructions) next to all of the air vents on the front, right and back sides?

Is there adequate water supply? Cycling power resets any silenced alarms. After turning the unit on,

wait 1 minute to see if ‘Water not filling properly’ alarm activate

This manual is related to the following products: