INSTALLATION
Unpacking
This product has been completely tested, cleaned and packed for shipment. Carefully remove all packing
material. Please examine the chamber completely. Should any damage be found, notify the delivering
carrier immediately. Report any shortages to your local distributor or contact CARON customer service at
740-373-6809, 800-648-3042 (USA only) or [email protected].
Chamber Location
The photostability chamber is an instrument and should be treated accordingly. The chamber must be
located in a dry, clean, and level area. Allow a 2 inch (51 mm) clearance from the back of chamber for
proper air circulation. Locate the chamber in an area out of direct sunlight and away from heating and
cooling ducts. Ambient temperature should be 18°C - 25°C and avoid locations where there are excessive
ambient air temperature fluctuations. The chamber can exhaust 4000 BTU/hr (1170 W) into the
surroundings. The chamber operates quietly around 60 dBA at 6 ft (2 m) away.
Caution: Do not put more than 10 lbs (4.5 kg) on top of unit.
The figure below shows a right view of the chamber with access port locations. Install rubber stoppers into
both access ports. Press in firmly.