Maintenance – Caron 6545 User Manual
Page 51
6540 Series Operations Manual Rev_I.docx
4/7/2014
51
MAINTENANCE
Electrical components (circuit breakers, fuses, relays, power supplies, etc.) are located on the electrical ‘sled’
or chassis. Access is through the front service panel. For additional service support, contact your local
distributor or CARON service department at
. When contacting CARON, be
sure to provide the following information from the information screen
Model number
Serial number
Electrical power
Software version
Display version
Routine Maintenance
Routine chamber maintenance is necessary to keep the chamber working properly. The recommended
chamber maintenance schedule is:
a) Replace lamps after 2000 hours of usage (not covered under warranty)
b) Vacuum or blow out condensing unit fins and PLC air vents every 500 hours (or more frequently
if dusty environment).
c) Clean interior chamber stainless steel with a non-abrasive glass cleaner.
d) Drain internal water reservoir (6545-1, 6545-2, 6545-3) if chamber is going to be unused for 10
days. See Information Humidity section of operations manual.
e) Apply electrical dielectric grease to connector pins on rear of both light banks
Required
information
WARNING: Before removing control panel, disconnect all power!