beautypg.com

Caron 6545 User Manual

Page 52

background image

6540 Series Operations Manual Rev_I.docx

4/7/2014

52

Brief Troubleshooting

Touch screen is blank

To maximize the touch screen optics, the touch screen will go into screen saver mode after being un-touched

for 30 minutes. To re-activate the touch screen (while the unit is still on), simply press anywhere on the

screen and the screen will illuminate. The power indicating light (illuminated blue ‘O’ in CARON) indicates

when the unit is on.

Chamber is alarming (see also alarm section of manual)

At the alarm screen, what alarm is indicated?

Was the alarm condition rectified?

Was the acknowledge alarm button pressed?

Has another alarm appeared in the alarm screen?

A ‘Lamps too hot’ alarm can’t be silenced until the lamps cool down to acceptable levels. To cool

down the lamps, turn unit off and open the door for 1 hour. If problem persists, contact service

personnel.

A ‘Lights dim’ alarm can also be reset by having the lights on at an intensity above the alarm level

for 1 minute.

Interior lights won’t come ‘on’

Are lights turned ‘on’?

Is light test ‘complete’?

Are any alarms activated? Silenced alarms are reset when unit is turned off.

Is door safety switch enabled?

Is correct light bank inserted?

Is light bank pushed in all the way?

Are the lamps inserted correctly?

Are all the lamps installed? If one lamp is not installed properly or burned out, it will affect the

performance of other lamps.

Interior lights have lower intensity than expected

Are all the lamps installed and on? If one lamp is not installed properly or burned out, it will affect

the performance of other lamps.

Are the lights dimmed?

Is there a light calibration factor applied?

Is something physically between the lamps and light detector?

Do the lamps need replaced?

Unit won’t reach temperature set point

Has the unit stabilized for 1 hour?

Is the temperature set point within the unit specification range?

Is the ambient temperature between 20ºC-25ºC?

If set point is above 25ºC, are the lights on?