beautypg.com

Caron 6545 User Manual

Page 60

background image

6540 Series Operations Manual Rev_I.docx

4/7/2014

60

APPENDIX A – VALIDATION (REFERENCE ONLY)

What is Validation/Qualification?

Validation/Qualification is the act of providing documented evidence that the installed equipment or systems

perform as specified during the design phase and meets the specifications of the manufacturer, the owner,

and applicable regulations. Typically, one would Qualify equipment but Validate a process. Steps include

Installation, Operation, and Performance Qualification. Validation/Qualification is more elaborate than

Equipment Manufacturer’s Installation in that it involves several additional steps farther performing system

calibration, worst case and functionality testing. Calibration of individual components is recommended prior

to validation.

Recommended Equipment for Validation/Qualification

Data acquisition system

Calibrated radiometer/photometer with calibrated UVA and VIS detectors

10 calibrated temperature sensors

2 calibrated humidity probes (if humidity control only)

Recommended validation steps to include

Controller functionality testing

System calibration of light controller (UVA and VIS light)

Light uniformity mapping

Chamber temperature mapping

Chamber humidity mapping (reference only)

Comprehensive final report

Light Detector Calibration

The UVA and VIS light detectors integrated into the chamber are high quality instruments specifically suited

for the application. These features include:

Detectors are calibrated to NIST traceable standards

Detectors are planar, cosine-corrected and have a Teflon hemisphere

UVA light detector calibrated to exact lamp’s spectral power distribution

VIS light detector spectral response closely follows CIE photopic action spectra & wide band width

For testing to a minimum exposure/dose level (such as ICH Q1B confirmatory testing), it may be useful to

calibrate the light detector to correspond with the minimum intense shelf location (see calibration).

Temperature and Humidity Calibration

To isolate the disturbance of lamp radiation, temperature and humidity sensors are mounted out of direct

lighting and measure air conditions entering the chamber. With the lights fully on, the chamber displayed

temperature (and humidity) may be different than a sensor on the shelf because of the light radiation. Any

temperature sensor in direct lighting should be wrapped in aluminum foil or similar material to measure air

temperature. A calibration offset can be applied to equate to conditions anywhere within the chamber (see

calibration).

This manual is related to the following products: