• Trouble Shooting
The DCR-1003, DCR-1004 and DST sensor are a rugged, reliable level
measurement system that is easy to install and setup. But occasionally problems
will occur during set up. A list of symptoms and possible corrective actions are
provided for troubleshooting.
SYMPTOM
CAUSE
ACTION
DISTANCE DISPLAY ON
Short circuit
• Verify coaxial connections
COM. LED IS OFF
on the transducer
and wiring on cable or connector
• Sensor failure
COM. LED FLASHING
Communication
• Check coaxial cable and
EVERY .5 SECONDS
error
connectors for tight connection
• Sensor failure in communication
circuit
COM. LED FLASHING
Loss of echo
• Poor target characteristics see
EVERY SECOND
“understanding ultrasonics”
section
• Sensor failure in transducer circuit
DISPLAY READS ----
Display overflow
• Change calibration, multiplier, or
decimal point position
• DST & DCR are incompatible,
upgrade DST software
DISPLAY WILL NOT
Sensor is seeing the
• Remove detected object at
CHANGE BUT LEVEL
wrong target
measured distance
DOES
• Mount the sensor away from tank
seams or obstructions
DISPLAY WILL ONLY
Sensor is receiving
• Check transducer installation for
READ AT CLOSE
strong echoes in its
smooth sound propagation into
DISTANCE
blanking distance
tank or to desired target
• Increase blanking distance mode 2