Example 9: Flume
A 1 meter wide cutthroat flume has been installed on the outlet of a water
treatment plant. Flow is to be displayed in liters/sec. (LPS).
The associated flow equation for the 1 meter flume is:
Flow in LPS = 11.48 * H^1.56
where H is the upstream head measurement as specified
for this flume type.
To program the DCR-1003, 1004 for this application, the following modes
must be set:
MODE
VALUE
DESCRIPTION
Operation
1
3
Set units to meters because the output is to be displayed in liters
per second.
3
4
Select Open Channel Monitoring for application
5
0.60 m
Do not set closer than the minimum blanking distance
recommended for the DST sensor type being used
Open Channel Flow
38
11
Constant ‘K’ above the decimal point
39
4800
Constant ‘K’ below the decimal point
40
.0
Decimal to one place
41
SPAN
Distance from .5 in. behind the transducer face to the bottom of
the flume using a tape measure
42
1
Exponent ‘X’ above the decimal point
43
5600
Exponent ‘X’ below the decimal point
.6 METERS
SPAN
REQUIRED BLANKING DIST.