beautypg.com

APG DCR-1004 user manual User Manual

Page 26

background image

DCR-1003 and DCR-1004

Rev. A3, 10/08

26

Automation Products Group, Inc.

APG...Providing tailored solutions for measurement applications

Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • [email protected]

VOLUME
A very popular use for the DCR-1003, 1004 is to do volumetric conversions

for the amount of product in a tank. If 0 is selected as the tank type , the zero
point is at the DST, and the display will show the distance to the product. If
tank type 1 - 5 are selected, the zero point is the bottom of the tank, and the
display will show how much product is in the tank.

MODE DESCRIPTION PARAMETERS

EXPLANATION

29

Tank Type

Range = 0 - 5

Selects the type of tank

0 - distance to level

to be monitored

1 - level indication

Set by NUM UP or NUM DN

2 - horizontal flat ends
3 - horizontal with
spherical ends.
4 - spherical tank
5 - conical bottom
Default = 0

Converting the output into units of volume other than (ft.

3

), (in.

3

) ,or (m

3

)

(determined in mode 1) is often preferred. To accomplish this, a conversion
factor needs to be entered into modes 30 and 31. For example, converting the
output from cubic inches to liters requires a conversion factor of 0.0164 (1 in.

3

= 0.0164 liters). If you are using tank type 1, you must multiply the area of
the bottom of the tank by the conversion factor and enter the result into modes
30 and 31. For example, if you have a flat-bottom tank (type1) with a diameter
of 110 in, you must first figure the area of the tank bottom. (Area of a circle = tt
x r

2

)

3.1415 x 55

2

= 9503.0375 IN

2

. The area is then multiplied by a conversion

factor. 9503.0375 x 0.0164 = 155.8498. This is the number you would enter
into modes 30 and 31. This is only necessary with tank type 1. If one of tank
mode 2-5 is selected, then the only the conversion factor needs to be entered in
modes 30 and 31. For example, to convert from (ft.

3

) to gallons will require a

conversion factor of 7.4805 (1 ft.

3

= 7.4805 gallons) to be entered into modes

30 and 31. Using the values entered in modes 30-35, along with the measured
distance to the level, the DCR calculates the volume in gallons.

30

Constant Above

Range = 0 - 9999

Sets the portion of the multiplier

Decimal Point

Default = 1

above the decimal point. Used
for converting between units of
volume. The relay and analog
outputs must be programmed in
the same units used here.

31

Constant Below

Range = .0000 - .9999

Sets the portion of the multiplier

This manual is related to the following products: