beautypg.com

APG DCR-1004 user manual User Manual

Page 24

background image

DCR-1003 and DCR-1004

Rev. A3, 10/08

24

Automation Products Group, Inc.

APG...Providing tailored solutions for measurement applications

Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • [email protected]

ANALOG
The analog output is only available on controller model DCR-1004. The

analog circuit comes standard with 4-20 mA, but may be configured for 0-20
mA. A hardware change will allow 0-5 V, or 0-10 V as the output.

The desired analog output should first be selected in mode 24. The two end

points must then be entered. For performing a distance to level measurement,
the zero point of distance will be at the transducer and the 4-20 limits will be
programmed in the same units as mode 1. For volumetric or open channel
monitoring, the zero point will be at the empty point of the tank or flume and
the end points must be set in the units used in the volume or flow measurement.
Keep this in mind when referring to the diagram below.

MODE DESCRIPTION

PARAMETERS

EXPLANATION

24

Analog Output

Range = 0 - 1

Selects the type of analog output

Select

0 = 4 to 20mA

output to be used

1 = 0 to 20mA

Set by NUM UP or NUM DN

Default = 0

25

0,4 mA Distance

Units = mode 1

Sets the minimum distance of

Default = 3.00 ft.

the analog output slope

26

20 mA Distance

Units = mode 1

Sets the maximum distance of

Default = 4.00 ft.

the analog output slope

27

0,4 mA Calibration

Fine tunes the minimum current
sourced on the analog output

28

20 mA Calibration

Fine tunes the maximum current
sourced on the analog output

This manual is related to the following products: