83

Voltage/frequency:

100–240 V ±10 %, 50–60 Hz

Fuses:

T0, 1.0 A – 5 x 20 mm (2 pcs.)

Power consumption:

60 W

Excess-voltage category:

II

Protection class:

I

Degree of contamination:

2

Voltage range
for the following modules:

Eukaryotic cells:

20– 100 V, in increments of

1 V

100–1,000 V, in increments of 10 V

1,000–1,200 V, in increments of 100 V
Interval between pulses: 1 minute

Cell fusion:

0,

5– 100 V, in increments of

1 V

100– 300 V, in increments of 10 V

Interval between pulses: 1 second

Alternating voltage: 0,

1.0–

10 V, in increments of 0.1 V

Bacteria:

200–1,000 V, in increments of 10 V

1,000–2,500 V, in increments of 100 V
No multiple pulse

Capacitor:

10 µF, 2,500 V impulse discharge

Resistance:

600

parallel

56

series

Charging time:

<30 seconds

No. of pulses (n):

0–99

Time constant

τ

of discharge:

Eukaryotic cells:

15–500 µs, in increments of 5 µs,

fading e-function

Bacteria:

Nominal 5 ms,
with an impedance of the sample of 3.3 k

Measurement and display from 0.8–6.0 ms

Duration of alignment:

0–95 s, in increments of 5 s

Duration t of pulse:

Cell fusion:

0, 15 µs–300 µs, in increments of 5 µs,

square-wave pulse

RS 232 interface:

9,600 baud, 8 bits, no parity,
2 stop bits, XON/XOFF

Ambient temperature:

max. 40 °C

Relative ambient humidity: max. 80 %

8 Technical data