8 technical data – Eppendorf Multiporator User Manual
Page 41
83
Voltage/frequency:
100–240 V ±10 %, 50–60 Hz
Fuses:
T0, 1.0 A – 5 x 20 mm (2 pcs.)
Power consumption:
60 W
Excess-voltage category:
II
Protection class:
I
Degree of contamination:
2
Voltage range
for the following modules:
Eukaryotic cells:
20– 100 V, in increments of
1 V
100–1,000 V, in increments of 10 V
1,000–1,200 V, in increments of 100 V
Interval between pulses: 1 minute
Cell fusion:
0,
5– 100 V, in increments of
1 V
100– 300 V, in increments of 10 V
Interval between pulses: 1 second
Alternating voltage: 0,
1.0–
10 V, in increments of 0.1 V
Bacteria:
200–1,000 V, in increments of 10 V
1,000–2,500 V, in increments of 100 V
No multiple pulse
Capacitor:
10 µF, 2,500 V impulse discharge
Resistance:
600
Ω
parallel
56
Ω
series
Charging time:
<30 seconds
No. of pulses (n):
0–99
Time constant
τ
of discharge:
Eukaryotic cells:
15–500 µs, in increments of 5 µs,
fading e-function
Bacteria:
Nominal 5 ms,
with an impedance of the sample of 3.3 k
Ω
Measurement and display from 0.8–6.0 ms
Duration of alignment:
0–95 s, in increments of 5 s
Duration t of pulse:
Cell fusion:
0, 15 µs–300 µs, in increments of 5 µs,
square-wave pulse
RS 232 interface:
9,600 baud, 8 bits, no parity,
2 stop bits, XON/XOFF
Ambient temperature:
max. 40 °C
Relative ambient humidity: max. 80 %
8 Technical data
Multiporator_Text_en.fm Seite 83 Freitag, 9. Juni 2006 8:05 08