76

Step 2 – Starting cell alignment

Terminate the
entering procedure.
The parameters
are effective
immediately and
do not have to be
stored.

t

: Duration of pulse,

Unit µs
(microseconds).
Can be set in
increments of 5 µs
between 0 and
15 µs–300 µs.

n: Number of pulses

0–99. Can be set in
one-pulse increments.

Trigger alignment
U'

~

by pressing

the START key.

The information

appears in

the fourth line of the
display; alternating
voltage is applied
for the duration
selected.

At the end of the
experiment, an acoustic
signal is emitted and an
initial display appears
with the parameters that
have been set and used.

In addition, the time that
has elapsed (in minutes
and seconds) since the
end of the experiment is
shown in the bottom line.
This display disappears
after 99 minutes or when
any key is pressed.

12.AUG.00

13:42

U'

~

5V

t

60s

U

0V

t

15s

n2

U

"

~

0V

t

30s

5 Mode of operation