6 error messages – Eppendorf Multiporator User Manual
Page 39

81
07
Short
τ
The capacitor
was completely
emptied; it is
uncertain whether
the
τ
has been
executed. The
limiting element
was the capacitor
and not the pulse
regulator.
Reduce the conductivity
of the solution used.
Check whether the
correct cuvette type has
been inserted.
08
Low Resistance The current in the
cuvette is too high
and the alternating
voltage cannot be
maintained.
Reduce the conductivity
of the solution used.
09
Timeout
Measuring
τ
Timeout for the
τ
-
measurement of
the bacteria
module.
Contact SERVICE.
Bac-
teria
mo-
dule
τ
a
= 0.8 ms,
U
a
is
considerably
lower than the
parameter U
which has been
set.
–
Sparks in the
cuvette.
–
The current
during
discharge
(pulse) was too
high.
–
Technical error.
–
Use a cuvette which
has a greater
distance between the
electrodes.
–
Reduce the
conductivity of the
solution.
–
Contact SERVICE.
Error
no.
Error message Cause
Solution
6 Error messages
Multiporator_Text_en.fm Seite 81 Freitag, 9. Juni 2006 8:05 08
- epMotion 96 (76 pages)
- epMotion 5070 (100 pages)
- epMotion 5075 (130 pages)
- Centrifuge 5427 R (64 pages)
- Centrifuge 5427 R (104 pages)
- White Paper 14 (8 pages)
- Rolling Cabinet (34 pages)
- Mastercycler nexus (142 pages)
- Mastercycler nexus (118 pages)
- Concentrator plus (New Design) (48 pages)
- Concentrator plus (43 pages)
- Easypet 3 (38 pages)
- Xplorer (74 pages)
- Xplorer Adjustment (26 pages)
- AF2200 Plate Reader (72 pages)
- AF2200 Plate Reader (78 pages)
- G0.5 µPlate (32 pages)
- BioSpectrometer basic (104 pages)
- BioSpectrometer kinetic (106 pages)
- BioSpectrometer fluorescence (102 pages)
- Micro Test Tubes (5 pages)
- Microplates (10 pages)
- PiezoXpert (34 pages)
- Eporator (38 pages)
- MiniSpin (25 pages)
- MiniSpin (20 pages)
- Centrifuge 5702 (32 pages)
- 5702 Centrifuge (27 pages)
- 5702 Centrifuge (32 pages)
- C5702 RH Centrifuge (32 pages)
- 5418 Centrifuge (80 pages)
- 5418 Centrifuge (48 pages)
- 5424 Centrifuge (71 pages)
- 5424 Centrifuge (44 pages)
- 5430 Centrifuge (88 pages)
- 5430 Centrifuge (130 pages)
- 5804 Centrifuge (95 pages)
- 5804 Centrifuge (127 pages)
- 5804 Centrifuge (129 pages)
- TransferMan4 r (102 pages)
- TransferMan4 m (96 pages)
- InjectMan 4 (100 pages)
- InjectMan NI 2 (60 pages)
- InjectMan NI 2 (16 pages)
- PatchMan NP 2 (53 pages)