beautypg.com

5 mode of operation – Eppendorf Multiporator User Manual

Page 36

background image

78

Step 2 – Starting cell fusion

Trigger the cell
fusion process
by Pressing the
START key.

"Charge" appears
briefly in the fourth
line of the display.
Charging
commences.

During the charging
procedure,
alternating voltage
U'

~

is applied for the

selected duration.
The alternating
voltage is
represented by the
symbol

~

(sine-wave

oscillation).

Discharge occurs
after the pre-
counted alignment
time (0–95
seconds). This is
indicated by a flash
in the fourth line of
the display.
If the number n
which has been
entered is n >1,
a further discharge
occurs each time
after one second.

12.AUG.00

13:42

U'

~

5V

t

30s

U

30V

t

15µs

n2

Charge

12.AUG.00

13:42

0:23

U'

~

5V

t

30s

U

30V

t

15µs

n2

5 Mode of operation

Multiporator_Text_en.fm Seite 78 Freitag, 9. Juni 2006 8:05 08