5 mode of operation – Eppendorf Multiporator User Manual
Page 30
72
Step 2 – Starting electroporation
–
Trigger the
electroporation
process by pressing
the START key.
–
"Charge" appears in
the third line of the
display. Charging
commences.
–
After the charging
procedure has
ended, discharging
occurs. This is
indicated by a flash
in the third line of
the display.
If more than one pulse
has been set, "Wait"
appears in the display,
accompanied by the
remaining waiting time.
After the experiment has
ended, a double acoustic
signal is emitted and the
initial information
appears
in the display with the
parameters which have
been used.
The time which has
elapsed since the end of
the experiment appears
in the bottom line in
minutes and seconds.
After 99 minutes or after
any key has been
pressed, this information
disappears from the
display.
12.AUG.00
13:37
U
500V
τ
220µs
Charge
12.AUG.00
13:37
U
500V
τ
220µs
Wait
0:57
12.AUG.00
13:37
U
500V
τ
220µs
n
2
READY SINCE 1:36
5 Mode of operation
Multiporator_Text_en.fm Seite 72 Freitag, 9. Juni 2006 8:05 08
- epMotion 96 (76 pages)
- epMotion 5070 (100 pages)
- epMotion 5075 (130 pages)
- Centrifuge 5427 R (104 pages)
- Centrifuge 5427 R (64 pages)
- White Paper 14 (8 pages)
- Rolling Cabinet (34 pages)
- Mastercycler nexus (118 pages)
- Mastercycler nexus (142 pages)
- Concentrator plus (New Design) (48 pages)
- Concentrator plus (43 pages)
- Easypet 3 (38 pages)
- Xplorer (74 pages)
- Xplorer Adjustment (26 pages)
- AF2200 Plate Reader (78 pages)
- AF2200 Plate Reader (72 pages)
- G0.5 µPlate (32 pages)
- BioSpectrometer basic (104 pages)
- BioSpectrometer kinetic (106 pages)
- BioSpectrometer fluorescence (102 pages)
- Micro Test Tubes (5 pages)
- Microplates (10 pages)
- PiezoXpert (34 pages)
- Eporator (38 pages)
- MiniSpin (20 pages)
- MiniSpin (25 pages)
- Centrifuge 5702 (32 pages)
- 5702 Centrifuge (27 pages)
- 5702 Centrifuge (32 pages)
- C5702 RH Centrifuge (32 pages)
- 5418 Centrifuge (80 pages)
- 5418 Centrifuge (48 pages)
- 5424 Centrifuge (71 pages)
- 5424 Centrifuge (44 pages)
- 5430 Centrifuge (88 pages)
- 5430 Centrifuge (130 pages)
- 5804 Centrifuge (127 pages)
- 5804 Centrifuge (129 pages)
- 5804 Centrifuge (95 pages)
- TransferMan4 r (102 pages)
- TransferMan4 m (96 pages)
- InjectMan 4 (100 pages)
- InjectMan NI 2 (60 pages)
- InjectMan NI 2 (16 pages)
- PatchMan NP 2 (53 pages)