4 device description – Eppendorf Multiporator User Manual

Page 22

background image

64

When screwing the core, make sure that no air bubbles form.

Fig. 10: Filling the Helix fusion chamber (schematic representation)

Leave the closed Helix fusion chamber upside-down on the coaxial
connector until the fusion process has been completed!

4.8.6 Fusion and cell suspension extraction

Slide the fusion insert firmly into the device.

Place the closed Helix fusion chamber onto the coaxial connector
and lock it into place by rotating it a quarter-turn.

Carry out the fusion with the set parameters (see chap. 5.3).

Loosen the Helix fusion chamber carefully by rotating it a quarter-turn
and then remove it.

To open the chamber, place it in a vertical position (e.g. in the stand).

Unscrew the core carefully from the beaker.

Using a small quantity of the appropriate medium, rinse off any cell
suspension that may be on the core.

The cell suspension from the beaker is mixed with the cell suspension
that has been rinsed off and is then treated further as required.