4 device description – Eppendorf Multiporator User Manual
Page 22
64
When screwing the core, make sure that no air bubbles form.
Fig. 10: Filling the Helix fusion chamber (schematic representation)
Leave the closed Helix fusion chamber upside-down on the coaxial
connector until the fusion process has been completed!
4.8.6 Fusion and cell suspension extraction
–
Slide the fusion insert firmly into the device.
–
Place the closed Helix fusion chamber onto the coaxial connector
and lock it into place by rotating it a quarter-turn.
–
Carry out the fusion with the set parameters (see chap. 5.3).
–
Loosen the Helix fusion chamber carefully by rotating it a quarter-turn
and then remove it.
–
To open the chamber, place it in a vertical position (e.g. in the stand).
–
Unscrew the core carefully from the beaker.
–
Using a small quantity of the appropriate medium, rinse off any cell
suspension that may be on the core.
–
The cell suspension from the beaker is mixed with the cell suspension
that has been rinsed off and is then treated further as required.