4 joystick, 5 software and parameters, 2 error messages – Eppendorf TransferMan4 r User Manual
Page 80: 1 warnings, Joystick, Software and parameters, Error messages 7.2.1, Warnings

80
Troubleshooting
TransferMan
®
4r
English (EN)
7.1.4
Joystick
7.1.5
Software and parameters
7.2
Error messages
7.2.1
Warnings
The display does not
show anything or the
device cannot be
activated although the
device is connected.
• The power cable or power
plug is loose.
• The device is switched off.
Check the power cable or
the power plug.
Switch the device on.
Symptom/message
Cause
Remedy
Outer movement range
of the joystick does not
work.
• Outer movement range is
deactivated.
In the
Installation
menu, set
the value for
Dyn-mode
to
ON
.
Symptom/message
Cause
Remedy
Parameters are not
accessible for certain
operations.
–
Enter parameters again.
Carry out a reset and reset
all parameters to the works
settings.
Readjust device.
Symptom/message
Cause
Remedy
WARNING 1
• FemtoJet is not connected.
Disconnect the data cable to
the FemtoJet and connect it
again.
Switch on the FemtoJet.
WARNING 3
• X-module is not connected.
Switch off the device.
Connect the plug of the
X-module to the control
board and tighten.
Switch the device on.
WARNING 4
• Y-module is not connected.
Switch off the device.
Connect the plug of the
Y-module to the control
board and tighten.
Switch the device on.
Symptom/message
Cause
Remedy