beautypg.com

4 joystick, 5 software and parameters, 2 error messages – Eppendorf TransferMan4 r User Manual

Page 80: 1 warnings, Joystick, Software and parameters, Error messages 7.2.1, Warnings

background image

80

Troubleshooting
TransferMan

®

4r

English (EN)

7.1.4

Joystick

7.1.5

Software and parameters

7.2

Error messages

7.2.1

Warnings

The display does not
show anything or the
device cannot be
activated although the
device is connected.

• The power cable or power

plug is loose.

• The device is switched off.

Check the power cable or
the power plug.

Switch the device on.

Symptom/message

Cause

Remedy

Outer movement range
of the joystick does not
work.

• Outer movement range is

deactivated.

In the

Installation

menu, set

the value for

Dyn-mode

to

ON

.

Symptom/message

Cause

Remedy

Parameters are not
accessible for certain
operations.

Enter parameters again.

Carry out a reset and reset
all parameters to the works
settings.

Readjust device.

Symptom/message

Cause

Remedy

WARNING 1

• FemtoJet is not connected.

Disconnect the data cable to
the FemtoJet and connect it
again.

Switch on the FemtoJet.

WARNING 3

• X-module is not connected.

Switch off the device.

Connect the plug of the
X-module to the control
board and tighten.

Switch the device on.

WARNING 4

• Y-module is not connected.

Switch off the device.

Connect the plug of the
Y-module to the control
board and tighten.

Switch the device on.

Symptom/message

Cause

Remedy