Operation
TransferMan
®
4r
English (EN)
6.8.1
Saving a position
Prerequisites
• An application has been selected.
6.8.2
Move to the position using the softkey
Prerequisites
• At least one position is stored.
1. Press a softkey with a stored position.
• You will move to the selected position.
• The joystick is deactivated until the position has been reached.
• The LEDs flash.
• In the softkey status field a filled circle is displayed.
1. Move the capillary to the required
position.
2. Hold the
Pos 1
softkey for approx. one
second to store the working position of
the capillary.
• An acoustic signal will sound.
•
Pos 1
is marked.
• The stored position is displayed in the
status field.
3. Move the capillary to the desired
position (e.g., parking position).
4. Press and hold the
Pos 2
softkey for
approx. one second to store the parking
position of the capillary.
• An acoustic signal will sound.
•
Pos 2
is marked.
• The stored position is displayed in the
status field.
If a stored position is lower than the lower limit (
Z-axis Limit
), the position on the
Z-axis will only be approached up to the defined limit.