2 executing the step injection function, 13 home function, 1 home menu and parameters – Eppendorf TransferMan4 r User Manual
Page 66: 2 setting the parameter for home, Executing the step injection function, Home function 6.13.1, Home menu and parameters, Setting the parameter for home

66
Operation
TransferMan
®
4r
English (EN)
6.12.2 Executing the
Step injection
function
Prerequisites
• A FemtoJet is connected.
•
Step injection
is assigned to a free softkey.
1. Activate
Step injection
.
2. Set
Step inj. dist.
3. Set the injection time on the FemtoJet.
4. Select
Step inj. dist.
5. Close the menu.
• The application screen displays
Step injection
.
6. Trigger the injection by selecting
Inject
on the FemtoJet.
• The function is executed.
6.13
Home
function
The
Home
function quickly moves the capillary out of the work area and is suitable for a
quick change of capillary.
6.13.1
Home
menu and parameters
Abb. 6-3:Menu –
Fig. 6-3:
Menu –
Home
6.13.2 Setting the parameter for
Home
1. Set the speed.
2. Set the offset.
3. Close the menu.
Parameter
Value
Range of values Increment Standard
Home speed
Set the speed of the Home
function
5 – 10000
5
7500
Home offset
Set vertical offset in μm
5 – 20000
5
0