English (EN)

6.13.6 Terminate the

home

function.

6.14

Clean

function

With this function you can move the capillary over a defined distance out of the medium.
In the process, outside contamination is removed from the capillary at the medium
boundary.

6.14.1

Clean

menu and parameters

Abb. 6-4:Menu –

Fig. 6-4:

Menu –

Clean

1. Press

Back manual

.

The

home

function is terminated.

2. Move the capillary manually using the

joystick.

Parameter

Value

Range of values Increment Standard

Home speed

Set the speed of the clean
function

5 – 10000

5

7500

Clean distance

Set distance in μm

0 – 20000

5

10000