beautypg.com

11 speed function, 1 speed menu and parameters, 2 setting the parameter for speed – Eppendorf TransferMan4 r User Manual

Page 64: Speed function 6.11.1, Speed menu and parameters, Setting the parameter for speed

background image

64

Operation
TransferMan

®

4r

English (EN)

6.11

Speed

function

In the

Speed

menu you can set the size of the working range and the speed at which

certain positions are approached or movements executed.

6.11.1

Speed

menu and parameters

Abb. 6-1:Menu –

Fig. 6-1:

Menu –

Speed

6.11.2 Setting the parameter for

Speed

1. Select the desired parameter.

2. Set the value for the parameter.

3. Close the menu.

Parameter

Value

Range of values Increment Standard

Coarse

Set size in μm

5 – 12500

5

6000

Fine

Set size in μm

5 – 2000

5

250

X-fine

Set size in μm

0 – 600

1

80

Step inj. speed

Set speed in μm per second

5 – 10000

5

300

Position speed

Set speed in μm per second

5 – 10000

5

1500

Home speed

Set speed in μm per second

5 – 10000

5

7500

If a value of 0 is set for the parameter

X-fine

, then the option of changing

between the

Fine

and

X-fine

working ranges is deactivated.