Operation
TransferMan
®
4r
English (EN)
6.12
Step injection
function
With this function you can execute a straight injection over a defined distance.
Step
injection
can only be executed when a FemtoJet is connected.
6.12.1
Step injection
menu and parameters
Abb. 6-2:Menu –
Fig. 6-2:
Menu –
Step injection
Parameter
Value
Range of values Increment Standard
Step injection
Switch function on or off
OFF/ON
–
OFF
Step inj. speed
Set inspection speed in μm per
second
5 – 10000
5
300
Step inj. dist.
Set injection path in μm
1 – 2000
1
20
Step inj. axial
Switch axial movement of the
capillary on or off
OFF/ON
–
On
Step inj. back
Switch backward movement of
the capillary on or off
OFF/ON
–
On
Synchr. inject.
Set the time of the injection
MOVE
IMMEDIATE
LIMIT
PRESSURE
–
LIMIT
Parameter
Range of values Function
Synchr. inject.
MOVE
Injection movement without injection pressure
IMMEDIATE
Injection pressure starts with injection movement
LIMIT
Injection pressure at the end of the injection distance
(
Step inj. dist.
)
PRESSURE
Injection pressure without injection movement