Technical data – IKA HBC 10 baisc User Manual
Page 38

38
Technical data
IC basic
HBC 5 basic
HBC 10 basic
Nominal voltage
VAC
230 ±10% (EURO) / 115 ±10% (USA)
Frequency
Hz
50/60
Max. input power
W
2650 (230 VAC)/1400 (115 VAC)
Working temperature range
(RT+10 at 1000rpm)
°C
RT + 10 ... 200
Operating temperature range
(with forced cooling)
°C
-20 ... 200
Temperature stability – Internal temperature
control 70°C, water (according to DIN12876) K
± 0,02
Temperature control
PID
Temperature measurement, absolute accuracy
Internal (int) (adjustable by calibration)
External (ext) (adjustable by calibration)
K
K
± 0,2
± 0,2
External Pt 100.5 temperature sensor tolerance
to DIN EN 60751 class A, ≤ ± (0.15 + 0.002 x |T|),
e.g. at max. 100°C (adjustable by calibration)
K
± 0,35 (at 100 °C)
Temperature setting
Knob
Temperature setting resolution
K
0,1
Temperature display
LED
Temperature display resolution
K
0,1
Classification according to DIN12876-1
Class III (FL) suitable for flammable and non-flammable fluids
Safety circuit (adjustable)
°C
0 ... 260
Safety temperature display
LED
Heating capacity
W
2500 (230 VAC)/1250 (115 VAC)
Pump speed (adjustable)
rpm
1000 ... 4000
Max. pump pressure/suction
bar
0,5/0,3
Max. flow rate (at 0 bar)
l/min
20
Bath volume
I
-/-
5 - 7
8 - 11
Sub-level protection
Yes
Interface
USB, RS 232
Permitted on-time
%
100
Protection class according to EN 60529
IP 21
Protection class
I
Excess voltage category
II
Contamination level
2
Permitted ambient temperature
°C
+5 ... +40
Permitted ambient humidity
%
80
Dimension (W x D x H)
mm
285 x 291 x 313
275 x 500 x 406
275 x 510 x 456
Weight
kg
8,5
17,0
18,0
Operation at a terrestrial altitude
m
max. 2000
Subject to technical changes!