Error codes – IKA HBC 10 baisc User Manual
Page 36

36
Error codes
Any malfunctions during operation will be identified by an error message on the display.
Proceed as follows in such cases:
F
Switch off device using the main switch at the back of the device
F
Carry out corrective measures
F
Restart device
Error code
Effect
Cause
Solution
Err 01
Pump off
Heating off
No External Pt 100 temperature
sensor
- Check this sensor
Err 02
Pump off
Heating off
Motor Over current (rate
current)
- Reduce pump motor speed
- Use fluid with lower viscosity
- Check if the pump impeller is blocked
Err 03
Pump off
Heating off
Motor Over current (Max
current)
- Reduce pump motor speed
- Use fluid with lower viscosity
- Check if the pump impeller is blocked
Err 04
Pump off
Heating off
Motor Hall signal missing
- Reduce pump motor speed
- Use fluid with lower viscosity
- Check if the pump impeller is blocked
Err 05
Pump off
Heating off
Too high liquid level
- Check the liquid level and buoyage
Err 06
Pump off
Heating off
Too low liquid level
- Check the liquid level and buoyage
Err 07
Pump off
Heating off
Too high voltage
- Check the mains power
Err 08
Pump off
Heating off
Too low voltage
- Check the mains power
Err 09
Pump off
Heating off
Device internal temperature is
too high
- Check the ambient temperature and let the device
cool down
Err 10
Pump off
Heating off
PC communication failure
- Check communication cable
Err 11
Pump off
Heating off
Temperature difference
between control sensor and
safety sensor is too much
- Check safety temperature circuit and bath fluid
Err 12
Pump off
Heating off
Safety temperature alarm
- Check the bath temperature measurement
Err 13
Pump off
Heating off
Heater switched off by safety
circuit
- Check safety temperature set value, fluid level
Err 14
Pump off
Heating off
Fan error
- Check the fan and clean the grids at the rear side
If the actions described fails to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department;
- Send the device for repair, including a short description of the fault.