Fig. 10
Fig. 11
Low Level
High Level
WARNING
• Fluids
No.
IKA
®
Designation
Operating temperature range
(°C)
Safety temperature
(°C)
Flash point
(°C)
0
EG19_N10.100 (H
2
0 + Glycole)
-10 ... 100
100
125
1
EG34_N20.100 (H
2
0 + Glycole)
-20 ... 100
100
125
2
EG40_N25.100 (H
2
0 + Glycole)
-25 ... 100
100
125
3
EG45_N30.100 (H
2
0 + Glycole)
-30 ... 100
100
125
4
SI_P20.275 (Silcone Oil)
20 ... 275
275
300
5
SI_N30.145 (Silicone Oil)
-30 ... 145
145
170
6
SY_N10.100 (Synthetic Oil)
-10 ... 100
100
125
7
Water*
5 ... 95
95
—
8
Customized
*
Note: Tap water may be unsuitable for operation because the calcium carbonate content may cause calcification.
High purity water (from icon exchangers) and distilled or bi-distilled water are unsuitable for operation due to corrosive
properties of these media. High purity water and distillates are suitable as a medium after adding 0,1 g soda (Na
2
CO
3
,
sodium carbonate) per liter of water.
Fig. 9
• Filling and draining
- Before filling the fluid into the bath, open the front
cover as indicated in following figure.
- Check and make sure that the drain valve is closed (the
stop position in clockwise direction, see Fig. 10).
Note: Please note information in chapter “Commis-
sioning”.
- Connect the mains plug and turn on the device with
mains switch (2).
- The low level warning message appear on the display.
- Open the lid (16) and fill fluid to the bath.
Note: Pay attention to the fluid
level information on the display:
- To drain the fluid from the bath, connect a hose to the
drain port and turn the drain valve in counter clockwise
direction with a straight screwdriver.
Drain valve
Drain port
Stop position