Menu settings
MODE
Maximum temperature (HI T)
Minimum temperature (LO T)
Maximum speed (HI R)
Minimum speed (LO R)
Fluid type (FLUI)
The type of temperature PID control arithmetic (AUTO)
External and internal temperature control (ET)
Alarm and key tone (BEEP)
Calibration (CALI)
Menu
• Menu structure
Default settings
A
B
C
Proportional coefficient of PID (Kp 1)
Integrate time of PID (Ti 1)
Differential time of PID (Td 1)
AUTO 1
AUTO 0
External sampling time (EuT 2)
Proportional coefficient of PID (Kp 2)
Integrate time of PID (Ti 2)
Differential time of PID (Td 2)
DF t
FASt 0
FASt 1
ET 0
ET 1
BEEP 0
BEEP 1
CALI 0
CALI 2
CALI 3
activated
-
-
200 °C
0 °C
4000 rpm
1000 rpm
0
-
15,0
5,00
5,0
activated
120 s
1,0
1,5
0,0
100,0
-
activated
-
activated
activated
-
-
• Basic guide to setting the menu options:
F
Enter the menu by pressing ”Menu” key (E).
F
Turn the left knob (B) to select a menu option.
F
Turn the right knob (C) to change the menu value settings.
F
Confirm the menu settings by pressing the right knob (C).
• Operating mode (MODE)
Mode A: After power on/power failure no automatic
restart of functions.
Mode B: After power on/power failure automatic re-
start of functions, depending on previous settings.
Mode C: Set values (set in A or B) cannot be changed.
After power on/power failure automatic restart of func-
tions, depending on previous settings.
• Maximum temperature (HI T)
The maximum adjustable value: 200° C. This value can
be limited additionally by the selected liquid (No., see
“Settings” fluid type).
Note: The maximum value is limited by the set safety
temperature.
• Minimum temperature (LO T)
The minimum adjustable value: 0° C
• Maximum speed (HI R)
The maximum adjustable value: 4000 rpm.