2 execution errors – BNC 645 User Manual
Page 97
97
-221 Settings conflict; turned off infinite burst to allow
immediate trigger source
An infinite burst count is only allowed when an external or bus
(software) trigger source is used.
-221 Settings conflict; infinite burst changed trigger source to
BUS
An infinite burst count is only allowed when an external or bus
(software) trigger source is used.
-221 Settings conflict; burst period increased to fit entire burst
The specified number of cycles in the BURS:NYNC command takes
priority over the burst period and the waveform generator increased
the burst period to accommodate the burst count or the waveform
frequency.
-221 Settings conflict; burst count reduced to fit entire burst
Since the burst period is currently at its maximum, the waveform
generator has adjusted the burst count to accommodate the
waveform frequency.
-221 Settings conflict; triggered burst not available for noise
Noise function is not allowed in the triggered burst mode.
-221 Settings conflict; amplitude units changed to Vpp due to
high-Z load
dBm is not available as unit when the output termination is set to
“high impedance”.
-221 Settings conflict; trigger output disabled by trigger external
When using the external trigger source, the Trig Out signal is
automatically disabled.
-221 Settings conflict; trigger output connector used by burst
gate
If you have enabled burst , and selected gated burst mode, the Trig
Out signal is automatically disabled.