BNC 645 User Manual
Page 104
104
-222 Data out of range; large period limits minimum pulse width
The minimum pulse width for pulse waves with period larger than 10
s is larger. Refer to the table of minimum pulse width in Section 4.5
for details.
-222 Data out of range; pulse edge time limited by width; value
clipped to…
The specified pulse edge time does not conform to the following
condition and is adjusted to accommodate the pulse width:
EdgeTime ≤ 0.625 x PulseWidth
-222 Data out of range; pulse edge time limited by duty cycle;
value clipped to…
The specified pulse edge time does not conform to the following
condition and is adjusted to accommodate the duty cycle:
EdgeTime ≤ 0.625 x Period x DutyCycle
-222 Data out of range; period; value clipped to…
The pulse period is out of range and is adjusted.
-222 Data out of range; frequency; value clipped to…
The frequency is out of range and is adjusted.
-222 Data out of range; user frequency; value clipped to upper
limit
The frequency is out of range and is adjusted to the maximum value
allowed due to selection of arbitrary waveform.
-222 Data out of range; ramp frequency; value clipped to upper
limit
The frequency is out of range and is adjusted to the maximum value
allowed due to selection of ramp waveform.
-222 Data out of range; pulse frequency; value clipped to upper
limit
The frequency is out of range and is adjusted to the maximum value