-- R --


*RCL {0
1234}

Recall the instrument state previously stored in the non-volatile storage
location.

Storage location 0 stores the waveform generator’s power-on state and
storage location 1 to 4 are empty when the instrument is shipped out from
the factory.

You can not recall a state from a storage location which is empty.

Storage location 0 is automatically overwritten when the power is cycled.


*RST
Reset the waveform generator to its factory default, independent of the
MEM:STAT:REC:AUTO command setting. This command does not affect any
stored instrument states, arbitrary waveforms, or the I/O settings. It also
enables the front-panel display if it has been turned off. This command will abort
a sweep or burst in progress.

-- S --


*SAV {0
1234}

Store the current instrument state to the specified non-volatile storage location.

If there is any instrument state stored in the specified location, it will be
overwritten.

You can store an instrument state into location 0 only from the remote
interface. It is not accessible from the front panel.

Storage location 0 is automatically overwritten when the power is cycled.

If you store the instrument state for an arbitrary waveform and then delete
it later, the waveform data is lost and the waveform generator is not able
to output the waveform.

When you store an instrument state, the selected function, frequency,
amplitude, dc offset, duty cycle, symmetry, as well as any modulation
parameters are saved also.


*TST?
Order a complete self-test of the waveform generator. Returns “0” when the test
runs successfully, or “1” when the test fails.