8 arbitrary waveform errors, Rbitrary, Aveform – BNC 645 User Manual
Page 110: Rrors

110
performing. Use the CAL:SEC:STAT ON command with correct
security code to unsecure the calibration memory.
703 Calibration error; secure code provided was invalid
Security code provided in the CAL:SEC:STAT ON command was
invalid.
706 Calibration error; provide value is out of range
The specified calibration value is out of range.
707 Calibration error; signal input is out of range
The analog-to-digital converter (ADC) verified that the signal applied
to the Modulation In connector on the rear panel is out of range.
707 Calibration error; cal edge time: rise time cal error
707 Calibration error; cal edge time: fall time cal error
707 calibration error; cal edge time: default value loaded
A failure in the rise time or fall time circuitry has prevented calibration
from going forward.
850 calibration error; setup is invalid
Invalid calibration setup number specified.
851 calibration error; setup is out of order
Calibration setup procedures must be performed in a specific
sequence in order to be valid.
7.8 Arbitrary Waveform Errors
770 Nonvolatile arb waveform memory corruption detected
A checksum error was found in the non-volatile memory used to store
arbitrary waveform. As a result, The arbitrary waveform cannot be
retrieved.
781 Not enough memory to store new arb waveform; use
DATA:DELETE