Model dc80r – NOVUS DC80L User Manual
Page 5
5
4. MODEL DC80R
The MODEL DC80R RTD Process Calibrator measures and simulates resistance
values of 7 distinct types of RTDs, either in °C or °F. Simultaneous measurement
and signal simulation is not provided.
The specifications are valid for the period of one year, at temperatures between
18~28 °C.
Resistance measurement and simulation specifications:
Range
Measurement accuracy
( 4 W ± Ω )
Simulate
Accuracy ( ± Ω )
Excitation
current (mA)
0.00 Ω ~ 400.00 Ω
0.1
0.15
0.1 ~ 0.5
0.1
0.5 ~ 3.0
400.0 Ω ~ 1500.0
Ω
0.5
0.5
0.05 ~ 0.8
1500.0 Ω ~
3200.0 Ω
1
1
0.05 ~ 0.4
2
The excitation current applies only to the simulation mode. The excitation current
is provided by the target instrument (OHM meter of RTD meter) connected to the
calibrator.
RTD measurement and simulation specifications:
Mode
Range
Accuracy ( °C )
Excitation
current (mA)
Input
4 W
Input
2 W / 3 W Output
Pt10 385
-200 ~ 800 °C /
-328 ~ 1472 °F
NOT SPECIFIED
0.1 ~ 3.0
Pt50 385
-200 ~ 800 °C /
-328 ~ 1472 °F
0.7
1.0
0.7
0.1 ~ 3.0
Pt100 385
-200 ~ 800 °C /
-328 ~ 1472 °F
0.33
0.5
0.33
0.1 ~ 3.0
Pt200 385
-200 ~ 250 °C /
-328 ~ 482 °F
0.2
0.3
0.2
0.1 ~ 3.0
250 ~ 630 °C /
482 ~ 1166 °F
0.8
1.6
0.8
Pt500 385
-200 ~ 500 °C /
-328 ~ 932 °F
0.3
0.6
0.3
0.05 ~ 0.8
500 ~ 630 °C /
932 ~ 1166 °F
0.4
0.9
0.4
Pt1000
385
-200 ~ 100 °C /-
328 ~ 212 °F
0.2
0.4
0.2
0.05 ~ 0.8
100 ~ 630 °C
/212 ~ 1166 °F
0.2
0.5
0.2
Pt100 JIS
-200 ~ 630 °C /
-328 ~ 1166 °F
0.3
0.5
0.3
0.1 ~ 3.0
The excitation current applies only to the simulate mode. The excitation current is
provided by the target instrument (OHM meter of RTD meter) connected to the
calibrator.