beautypg.com

NOVUS RHT-P10-485-LCD User Manual

1 introduction, 2 specifications

background image

NOVUS AUTOMATION

1/6

Transmitters RHT-WM-485-LCD,

RHT-DM-485-LCD and RHT-P10-485-LCD

TEMPERATURE AND HUMIDITY TRANSMITTER –

INSTRUCTIONS MANUAL V1.0x B

1 INTRODUCTION

The RHT-WM-485-LCD, RHT-DM-485-LCD and RHT-P10-485-LCD

transmitters include high precision and stability sensors for measuring

temperature and relative humidity. They are microprocessed devices and,

therefore, allow being entirely configured through a RS485 interface, using

the Modbus RTU commands. The DigiConfig allows the configuration of all

transmitter features as well as its diagnostic procedure.
The transmitter can be configured or displaying the values of measured

temperature and Relative Humidity or the values of measured temperature

and Dew Point.

1.1 Identification
The identification label is fixed on the transmitter body. Check if the features

described on this label are in accordance with your order.
The RHT-WM-485-LCD and RHT-P10-485-LCD models intended for wall

mounting, while the RHT-DM-485-LCD model comes with a probe tip rod for

installation in ducts or through walls.
The following elements are located at the front side of the transmitter:

Figure 1 – Transmitter front

Communication Indicator (COM): Indicates when the device is receiving

data from the Modbus network.
Button

(Programming): Button used for configuration of the

communication parameters and for screen changing.
Temperature: Value of measured temperature.
Relative Humidity: Value of measured relative humidity.

Figure 2 – Second transmitter screen

Communication Address: Value programmed for identifying the transmitter

of the Modbus network. Programmable between 1 and 247.





2 SPECIFICATIONS

Operating temperature of

the transmitter

From 0 °C to 70 °C

Sensor and Rod (RHT-DM): See Figure 3.

Electromagnetic

compatibility

EM 61326:2000

Sensor measuring

range

Temperature: –40.0 °C to 100.0 °C.
Relative Humidity (RH): 0.0 to 100.0 % RH.

(No Condensing)

Dew Point: –40.0 °C and 100.0 °C

Measurement accuracy

See Figure 3.

Note: a small measurement error can be

eliminated adjusting the OFFSET parameter of

the software.

Measuring resolution

Temperature: 0.1 °C.

14 bits (16383 levels)

Relative Humidity (RH): 0.1 %.

12 bits (4095 levels)

Response time

Temperature: up to 30s in slow air movement.
Humidity: up to 8s in smooth air movement

(20 to 80 % RH).

Sampling interval

3 seconds

Power supply

12 Vdc to 30 Vdc, consumption < 16 mA

Case

Polycarbonate

Protection class

Product appropriate for applications which

require protection class up to IP65.

Electronic module case: IP65; Sensor

capsule: IP40

Dimensions

60 x 70 x 35 mm

DigiConfig software

operation environment

Configuration Software for Windows 98, NT,

2000, XP and Vista.

Menus in Portuguese, English or Spanish.

Configures, reads and exhibits data on the

screen.








This manual is related to the following products: