Maintenance – NOVUS DC80L User Manual

Page 13

background image

13

b) Simulating a 2-wireTransmitter (sink transmitter)

1. Press the power switch to

turn on the Calibrator.

2. Press the INPUT / OUTPUT

mode key to select

OUTPUT mode

.

3. Press the mA / mA % mode key

to select mA or mA % .

4. Press the increment

/

decrement keys

to set the desired output value.

5. Plug the red test lead into the V

jack

and the black one to

the COM jack .

6. Connect the test leads such that

the current enters the red lead

and returns to the power supply

through the black lead.

7. To change the output value,

press the increment / decrement

keys

and / or

the mA / mA % mode key .

Accessories: a pair of Test Leads

6. MAINTENANCE

Cleaning
Periodically wipe the case with a damp cloth and detergent; do not use abrasives

or solvents.

Calibration
Calibrate your calibrator once a year to ensure its performance according to the

specifications.

Replacing the Battery
Change the battery when the display signs

.

Turn off the power, unscrew the battery cover, and replace the batteries.