DC mV and V generation

1. Press the power switch to

turn on the Calibrator.

2. Press the INPUT / OUTPUT

mode key to select

OUTPUT mode .

3. Press the V / mV mode key

to select the desired output

scale. The mode indication (V or

mV) is shown in .

4. Use the increment / decrement

keys

to adjust

the output to the desired value.

5. Plug the red test lead in the “V

jack

and the black one to

the COM jack .

6. Connect the positive test lead

(red) and the negative test lead

(black) to the target equipment.

7. For changing the output value,

use the increment / decrement

keys

and the V

/ mV mode key until the

desired value is displayed.

DC mA measurement
a) Using external power supply

1. Press the power switch to turn

on the Calibrator.

2. Press the INPUT / OUTPUT

selection key

to select input

mode. The INPUT sign is shown

on the display .

3. Press the mA / mA % mode key

, to select mA or mA % .

4. Plug the red test lead in the mA

jack

and the black one to

the COM jack .

5. Connect the test lead according

to the signal polarity (red to the

positive, black to the negative).

6. The measurement is shown on

the display .