beautypg.com

9 special care, 1 care with sensors, 2 battery care – NOVUS RHT-Air User Manual

Page 8: 10 warranty

background image

RHT-Air

NOVUS AUTOMATION

8/8

9 SPECIAL CARE

The equipment needs special care since it is electronic:

• When opening it for fixation, you should keep away from the electronic

circuit due to the risk of damages caused by static electricity.

• The wiring should be carefully observed.
• When closing the box, the cover must be replaced properly, thus

ensuring that it is sealed.

9.1 CARE WITH SENSORS
The calibration of the humidity sensor can be altered if it is exposed to

contaminating vapors or extreme humidity or temperature conditions for

longer periods. In order to speed up the calibration, do as follows:

• Remove the sensor from the capsule.
• If solid particles are deposed on the sensor, wash it with water.
• Place the sensor in an oven at 80 °C (+ -10 °C) for 24 hours.
• Place the sensor for 48 hours in a place at around 20 and 30 °C and

humidity higher than 75% RH.

• Replace the sensor into the capsule.

9.2 BATTERY CARE

The Lithium battery, due to its chemical behavior, may present some

passivation symptoms when not used for long periods of time. This can be

observed by checking the battery voltage after 2 hours of equipment

operation. If the battery voltage measures less than 3.5 V, then the battery

may be passivated (not necessarily worn out). If this is the case, please try

applying the following procedures to recover the battery.

• Operate the equipment at its maximum power (14 dBm) and minimum

sample period (15 s).

• Let the instrument operate continuously in the above conditions for 24 h

while monitoring the battery voltage through reading register number 29

or the Diagnostic menu of the DigiConfig software.

• If the battery voltage exceeds 3.5 V, the battery can be considered

rehabilitated (depassivated). The equipment is then ready to use.

• In case the above doesn’t work, please remove the battery from the

equipment and discharge it for 5 minutes connecting a 220 ohm resistor

across the terminals.

• Then, reinstall the battery in the equipment and let it operate for 2 hours.

If the battery voltage remains above 3.5 V, then the depassivation

procedure was successful.

• If after going through the above procedures the battery still fails to work

properly, please contact Novus technical assistance for arranging battery

replacement.

IMPORTANT

The sensor used in this equipment may be damaged or de-calibrated if

exposed to contaminated atmospheres with chemical agents. Hydrochloric

acid, nitric acid, sulfuric acid and ammonia in high concentrations may

damage the sensor. Acetone, ethanol and propylene glycol may cause

reversible measurement errors.

10 WARRANTY

Warranty conditions are available on our web site www.novusautomation.com.