Campbell Scientific OBS500 Smart Turbidity Meter with ClearSensor Technology User Manual
Page 40

OBS500 Smart Turbidity Meter with ClearSensor™ Technology
Note that Edlog only allocates one input location for Instruction 105. Two
input locations are required—one for the pressure measurement and one for the
temperature measurement. The additional input location needs to be inserted
manually using the Input Location Editor. To get into the Input Location
Editor, select Edit/Input Labels or press the F5 key. Once in the Input Location
Editor, do the following steps:
1. Choose Edit/Insert Block.
2. After the Insert Block dialog box appears, type in a base name for the
input locations. Each input location will have the base name with an
underscore and a consecutive number.
3. In the Start Address field, type in the number of the first input location.
4. In the Number of InLocs field, type in 2 and select OK.
7.7 Operation in High Sediment Loads and Sandy Sediments
Sites with high sediment loads and large sand grains can be problematic for the
shutter and its motor. The recommendations provided in this section should
help reduce these problems.
Typically sites with high biological growth have relatively low
sediment loads.
1. Run the OBS500 in a normally open mode. For example, close then open
the wiper once every four hours. This reduces the wear on the motor
significantly, and save power. The interval can be adjusted over time.
Increase the interval if experiencing fouling. If the windows are staying
clean, slow it down even more. Example CRBasic programs are provided
at Appendix B.4, Examples for High Sediment Loads.
a.
M3! opens the wiper
b.
M4!, M5!, or M6! perform measurements when the wiper is open
c.
M7! closes the wiper
2. Clean the shutter assembly. The frequency that the shutter should be
cleaned depends on the sediment load and can vary from weeks to months
(step 3 can help you determine the required frequency for cleaning). Two
levels of cleaning should be done;
a. flush the wiper as it opens and closes with a stream of clean water, or
b. remove the wiper from the OBS500 by removing one screw and
follow the directions provided in Section 7.7.1, Wiper Removal
Procedure. Flush and clean.
3. Store the current used to open and close the slider. The open and close
SDI-12 instructions (M3! and M7!) output the current. Normally the
current is around 100 mA. As sand grits lodge in the groves, the resistance
to movement increases and the motor has to work harder. This increases
NOTE
30