OBS500 Smart Turbidity Meter with ClearSensor™ Technology

TABLE 7-7. SDI-12 and RS-232 Measurement Commands

Commands Process

Values Returned

aM7!

aC7!

Close Wiper

Send Data

Close wiper position count

Close max current count

Close timeout count

Close current (ma)

aM8!

aC8!

Send Wiper Data

Open close total count

Open wiper position count

Open max current count

Open timeout count

Close wiper position count

Close max current count

Close timeout count

aC9!

Open Wiper

Measure 100 Times

Close

Send Data

obs median

obs mean

obs standard deviation

obs min

obs max

ss median

ss mean

ss standard deviation

ss min

ss max

With the SDI-12 concurrent measurements (

aCx!), the datalogger

does not request the data until the next interval hits. For example,

if you have a 30-minute interval, you will not see the data for 30

minutes. There is not an equivalent

M command to the aC9!

command since the

M command is limited to nine returned values.

As the measurement data is transferred between the probe and datalogger

digitally, there are no offset errors incurred with increasing cable length as seen

with analog sensors. However, with increasing cable length, there is still a

point when the digital communications will break down, resulting in either no

response or excessive SDI-12 retries and incorrect data due to noise problems.

Using SDI-12 commands which add a CRC check (e.g.,

aMC!), can

significantly improve incorrect data issues.

7.4.1.1 SDI-12 Addresses

The OBS500 SDI-12 address can be set to 0 to 9, A to Z, or a to z which allows

multiple sensors to be connected to a single digital I/O channel (control port) of

an SDI-12 datalogger. (Most Campbell Scientific dataloggers support SDI-12.)

The OBS500 is shipped from the factory with the address set to 0. When it is

necessary to measure more than one OBS500, it is easiest to use a different

control port for each OBS500 instead of changing the address. If additional

control ports are not available, the address will need to be changed.

The address on the OBS500 can be changed by sending the SDI-12 change

address command

aAb!. The change address command can be issued from

most SDI-12 recorders. For example, to change the address of a sensor that has

a default address of 0 to the address of 1 the following command can be sent:

0A1!