OBS500 Smart Turbidity Meter with ClearSensor™ Technology

universal response curve. However, there is residual nonlinearity that is

removed by calibration and by computation of a TU value with a 2nd-order

polynomial. This section explains how to do a turbidity calibration.

0

1000

2000

3000

4000

Turbidity (TU)

0.0

0.2

0.4

0.6

0.8

1.0

N

or

m

al

iz

ed

O

B

S-

3+

R

es

po

ns

e

(O

PV

33

0

VC

SE

L)

FIGURE 7-5. Normalized response of OBS500 to AMCO Clear

®

turbidity. The inset shows the response function of a turbidity

sensor to high-sediment concentrations.

AMCO Clear

®

SDVB turbidity standards are used to calibrate an OBS500

sensor. SDVB standards are made for individual instruments. Standards made

for one model of turbidity meter cannot be used to calibrate a different model.

0

20000

40000

60000

SSC (mg/l)

0

1000

2000

3000

4000

Tu

rb

id

ity

(N

TU

)

A B

C

23