2 sdi-12 transparent mode, Sdi-12 transparent mode, 4. terminal emulator – Campbell Scientific OBS500 Smart Turbidity Meter with ClearSensor Technology User Manual

Page 31

background image

OBS500 Smart Turbidity Meter with ClearSensor™ Technology

The address may also be changed by connecting to the probe in the Device

Configuration Utility. Once connected, in the Settings Editor tab click in the

address box and enter the new address. Press Apply to save the changes.

7.4.1.2 SDI-12 Transparent Mode

The transparent mode allows direct communication with the OBS500. This

may require waiting for programmed datalogger commands to finish before

sending responses. While in the transparent mode, datalogger programs may

not execute. Datalogger security may need to be unlocked before the

transparent mode can be activated.

The transparent mode is entered while the PC is in telecommunications with

the datalogger through a terminal emulator program. It is most easily accessed

through Campbell Scientific datalogger support software, but it is also

accessible with terminal emulator programs such as Windows Hyperterminal.

To enter the SDI-12 transparent mode, enter the terminal emulator from

LoggerNet, PC400, or PC200W datalogger support software. A terminal

emulator screen is displayed. Click the

Open Terminal button.

For CR800 series, CR1000, and CR3000 dataloggers, press until the

datalogger responds with the prompt (e.g., “CR1000>” for the CR1000). Type

SDI12 at the prompt and press . In response, the query Enter Cx Port

1,3,5 or 7 will appear. Enter an integer value indicating the control port to

which the OBS500 is connected. A response of Entering SDI12 Terminal

indicates that SDI-12 Transparent Mode is active. Any of the SDI-12

commands may be entered (e.g.,

aM1! where a refers to the address). After

entering a command, the results may be viewed by entering

aR!.

FIGURE 7-4. Terminal Emulator

21