beautypg.com

Campbell Scientific CS135 Lidar Ceilometer User Manual

Page 54

background image

CS135 Ceilometer

46

CR LF = Carriage Return + Line Feed


LINE 3
scale9res9n9 energy9lt9tr9ti_bl9L0112HN159sum CR LF

where

scale (5 characters) = Scale parameter,%, 0 ... 99999. 100% is typical

res (2 characters) = Backscatter profile resolution in metres.

n (4 characters) = Profile length 385, 770, 1400, or 1500 samples.

energy (3 characters) = Laser pulse energy, %.

lt (3 characters including leading +/-) = Laser temperature, degrees C

tr (3 characters) = Window transmission, %

ti (2 characters) = Total tilt angle, degrees

bl (4 characters) = Background light, millivolts at internal ADC input (0 ... 2500)

"L0112HN15" (9 characters) = Reserved, defaults shown

sum = Sum of detected and normalized backscatter, 0 ... 999. Not used in Operating System

1

CR LF

= Carriage Return + Line Feed

This line is omitted from message 105.


LINE 4
SSSSSSSSSSSSSSSSSSSSSSSSSSSSS........(5 x 770 bytes) CR LF

The two-way attenuated normalised backscatter profile
(100000·srad·km)

-1

unless otherwise scaled by the Attenuated_SCALE parameter. Each sample is

coded with a 20-bit HEX ASCII character set; msb nibble and bit first, 2's complement.

The profile is not corrected for the tilt angle.

CR LF

= Carriage Return + Line Feed

This line is omitted from message 105.

NOTE

NOTE