45

Description of the most significant alarm word (left word, bits going left to right)
Bit Description
8000 XXXX XXXX

Transmitter shut-off

4000 XXXX XXXX

Transmitter failure

2000 XXXX XXXX

Receiver failure

1000 XXXX XXXX

Reserved for future use

0800 XXXX XXXX

Reserved for future use

0400 XXXX XXXX

Memory error

0200 XXXX XXXX

Light path obstruction

0100 XXXX XXXX

Receiver saturation

0080 XXXX XXXX

Reserved for future use

0040 XXXX XXXX

Reserved for future use

0020 XXXX XXXX

Reserved for future use

0010 XXXX XXXX

Reserved for future use

0008 XXXX XXXX

Reserved for future use

0004 XXXX XXXX

Reserved for future use

0002 XXXX XXXX

Reserved for future use

0001 XXXX XXXX

Ceilometer engine board failure

Description of the middle alarm word (middle word, bits going left to right)
Bit Description
XXXX 8000 XXXX

Window contamination

XXXX 4000 XXXX

Battery voltage low

XXXX 2000 XXXX

Transmitter expires

XXXX 1000 XXXX

High humidity

XXXX 0800 XXXX

Reserved for future use

XXXX 0400 XXXX

Blower failure

XXXX 0200 XXXX

Reserved for future use

XXXX 0100 XXXX

Humidity sensor failure

XXXX 0080 XXXX

Heater fault

XXXX 0040 XXXX

High background radiance

XXXX 0020 XXXX

Ceilometer engine board failure

XXXX 0010 XXXX

Reserved for future use

XXXX 0008 XXXX

Laser monitor failure

XXXX 0004 XXXX

Receiver warning

XXXX 0002 XXXX

Tilt beyond limit set by user, default 45 degrees

XXXX 0001 XXXX

Reserved for future use

Description of the least significant alarm word (right word, bits going left to right)
Bit Description
XXXX XXXX 8000

Blower is on

XXXX XXXX 4000

Blower heater is on

XXXX XXXX 2000

Internal heater is on

XXXX XXXX 1000

Working from battery

XXXX XXXX 0800

Reserved for future use

XXXX XXXX 0400

Self test in progress

XXXX XXXX 0200

Reserved for future use

XXXX XXXX 0100

Reserved for future use

XXXX XXXX 0080

Units are metres if on, else feet

XXXX XXXX 0040

Reserved for future use

XXXX XXXX 0020

Polling mode is on

XXXX XXXX 0010

Reserved for future use

XXXX XXXX 0008

Reserved for future use

XXXX XXXX 0004

Reserved for future use

XXXX XXXX 0002

Reserved for future use

XXXX XXXX 0001

Reserved for future use