beautypg.com

3 cl31 messages – Campbell Scientific CS135 Lidar Ceilometer User Manual

Page 50

background image

CS135 Ceilometer

42


CR LF

= Carriage Return + Line Feed



LINE 4
scale9res9n9 energy9lt9tr9ti_bl9pulse

9

rate9sum CR LF


where

scale (5 characters) = Scale parameter,%, 0 ... 99999. 100% is default

res (2 characters) = Backscatter profile resolution in metres.

n (4 characters) = Profile length 2048

energy (3 characters) = Laser pulse energy, %.

lt (3 characters including leading +/-) = Laser temperature, degrees C

ti (2 characters) = Total tilt angle, degrees


bl (4 characters) = Background light, millivolts at internal ADC input (0 ... 2500)

pulse (4 characters) = Pulse quantity x 1000 (0000-9999)

rate (2 characters) =
Sample rate, MHz, (00-99)

sum
= Sum of detected and normalized backscatter, 0 ... 999. Multiplied by scaling factor times
10

4

. At scaling factor 100 the SUM range 0 ... 999 corresponds to integrated backscatter 0 ... 0.

srad

-1.


CR LF

= Carriage Return + Line Feed



LINE 5
SSSSSSSSSSSSSSSSSSSSSSSSSSSSS........(2048 x 5 bytes) CR LF

The two-way attenuated normalised backscatter profile
(100000·srad·km)

-1

scaled by the Attenuated_SCALE parameter. Each sample is coded with a

20-bit HEX ASCII character set; msb nibble and bit first, 2's complement.

The profile is not corrected for the tilt angle.

CR LF

= Carriage Return + Line Feed

NOTE